Details
|
Technical data
|
|
Principle of measurements
|
Electrochemical Sensor, alcohol-specific
|
|
Measurement Range (BrAC)
|
0.000mg/L ~ 1.500 mg/L
|
|
Resolution
|
0.0001mg/L
|
|
Accuracy:
Max.Measuring error, in relation to the ethanol standard
|
Ethanol Vapor Concentration
C /(mg/L)
|
Deviation
|
|
Range 0 to 1‰
|
±0.05‰ absolute
|
|
Range 1 to 2‰
|
±5‰ of measured value
|
|
Range > 2‰
|
±5‰ of measured value
|
|
Calibration
|
Every 6 months
|
|
Temperature range-operation
|
-15c ~ 45…
|
|
Temperature range-storage
|
-30… ~ 70…
|
|
Blowing Time
|
3 seconds (breath flow :≥ 20L/min)
|
|
Warm-up Time
|
6 seconds after switching on
|
|
Recovery Time
|
5 seconds to 5 minutes, depends on Ethanol Vapor Concentration
|
|
Analysis Time
|
Less than 10 seconds
|
|
Optimum Storage Conditions
|
Temperature 25… ±5, Humidity: <80%RH,Pressure: 86kPa~106kPa, No vapor or electromagnetic disturbances in the environment are influencing the normal working conditions of the device.
|
|
Blowing time
|
3 seconds (breath flow: ≥20L/min)
|
|
Display
|
2.4 inch 320*240 dot matrix touch screen
|
|
Memory Capacity
|
Stores 4000 measured values with date and time
|
|
Communication Method for Printer
|
Bluetooth
|
|
Printer
|
External connected Bluetooth, Thermal or Dot Matrix Printer
|
|
Dimensions (L*W*T)
|
Approx. 140mm*60mm*24.5mm
|
|
Number of measurements per Battery Pack
|
More than 500 times (under ambient conditions)
|
|
Operation Voltage
|
State of charging
|
State of Operation
|
|
DC 5V
|
DC 3.3 ~ 4.2V
|
|
Batteries
|
1200mAh/10V polymerized battery
|
|
Weight
|
Approx. 180g (including printer, batteries, paper roll)
|